When collecting full spectrum data from any of Vernier’s Go Direct spectrophotometers listed below, the software will report more values when connected by USB than when connected by Bluetooth. This is done intentionally to optimize Bluetooth performance. The resolution for both connection types will still fall within the stated specifications because the spacing between reported wavelength values is smaller than the optical resolution of the spectrometers regardless of connection.
Go Direct® SpectroVis® Plus Spectrophotometer (GDX-SVISPL)
Go Direct® Visible Spectrophotometer (GDX-SPEC-VIS)
Go Direct® UV-VIS Spectrophotometer (GDX-SPEC-UV)
Go Direct® Fluorescence/UV-VIS Spectrophotometer (GDX-SPEC-FUV)
Go Direct® Emissions Spectrometer (GDX-SPEC-EM)